7 Troubleshooting
CyFlow Cube 6 V2m | Operating Manual | February 2020
33
Fault
Remedy
The result of the Calibration
Beads measurement is not
in the preselected regions of
the original configuration
script
Check within the software if you have loaded the correct
configuration file for your measurement.
Shake the bottle vigorously (e.g. by vortexing) and repeat
the measurement.
Check the date of expiry of the Calibration Beads.
Check the speed settings.
Perform a priming procedure.
Perform a cleaning procedure.
7.1.3 Measurements and data acquisition
Fault
Remedy
The device emits an
unusual / noisy sound after
starting the measurement
Check that the Waste bottle is closed properly and that
the bottle is not cracked.
Check all visible tubing and make sure that they are not
pinched.
Check that liquid (Sheath Fluid and sample) is dropping
into the Waste bottle after starting the measurement.
There is no data acquisition
visible (peaks/dots in
histogram/dot plot) during
the RUN phase
Check within the software if the laser/s is/are switched
on.
Check if the right trigger parameter has been selected
(choose FSC trigger as default setting, select SSC for
very small cells, e.g. bacteria).
Check gain values and the threshold.
Increase the flow rate.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SYSMEX CyFlow Cube 6 V2m and is the answer not in the manual?

SYSMEX CyFlow Cube 6 V2m Specifications

General IconGeneral
BrandSYSMEX
ModelCyFlow Cube 6 V2m
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals