6 Operation
26
CyFlow™ Cube 6 V2m | Operating Manual | February 2020
6.1.4 Performing a quality check
Calibration Beads
A quality check (QC) procedure is available for the device. This process controls scatter
and fluoresncence parameters as well as optical alignment of the device.
Requirements
Sample tube with Calibration Beads 3 µm, ready-to-use
Priming has been performed
Procedure
Figure 7: Selecting QC Mode
1. Open the QC Cube 6 V2m configuration file. [Prime Mode] is default.
2. Select the [QC Mode].
3. Press [Start].
4. Follow the instructions until the quality check is completed.