10 Technical data
42
CyFlow™ Cube 6 V2m | Operating Manual | February 2020
10.4 Fluidics
Item
Specification
Fluid system
Completely closed system for Sheath Fluid and sample
volumes
No fluid droplets or aerosols generated nor released from
the device
The pressure of the closed system is below 500 mbar
Flow cuvette
Synthetic quartz flow cuvette with centric flow channel
(capillary diameter 350 μm × 200 μm) for laminar sample
transport with Sheath Fluid
Sample delivery
Computer controlled precision syringe pump for
contamination-free sample transport.
Built-in vacuum pump for Waste bottle; vacuum pressure
is adjustable (computer controlled)
Sample volume
Continuous up to 1200 μl
Minimal content of sample volume 850 μl, an additional
dead volume of 50 μl is recommended for best results
200 μl for electrode-based precision absolute counting
(other counting volumes upon request)
Minimal sample volume 10 μl
10 μl to 1000 μl for syringe-based precision absolute
counting
Flow rates
Sample volume speed adjustable continuously between
0.1 μl/s and 19,9 μl/s)
Fluids volume
2 glass bottles (each 1 litre) for Sheath Fluid and waste
Bio safety system
Avoids sample droplets and sample cross contamination
(computer controlled)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SYSMEX CyFlow Cube 6 V2m and is the answer not in the manual?

SYSMEX CyFlow Cube 6 V2m Specifications

General IconGeneral
BrandSYSMEX
ModelCyFlow Cube 6 V2m
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals