5 Installation
5.1 Positioning the device
Carefully select the place where to position and operate the device, it should not be
moved without disconnecting its components. Comply with the following criteria when
positioning the device:
Place the device on a solid, dry and clean horizontal surface.
Place the device and the Sheath Fluid and Waste bottles on surfaces of identical
height, e.g. the same table, at a distance corresponding to the length of the sensor
cables and tubing.
Avoid smoke, dust, vibrations, direct sunlight and any source of direct heat.
Leave a minimum distance of 15 cm between the back of the device and a wall to
keep the ventilation effective.
Do not block the ventilation grid at the back of the device.
Do not place any objects on the device.
5.2 Installation and uninstallation
Installation, connection, disconnection and uninstallation may only be performed by
Authorised service personnel and in conformity with the applicable national rules and
regulations.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SYSMEX CyFlow Cube 6 V2m and is the answer not in the manual?

SYSMEX CyFlow Cube 6 V2m Specifications

General IconGeneral
BrandSYSMEX
ModelCyFlow Cube 6 V2m
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals