EasyManua.ls Logo

Tektronix CSA7404B - Page 195

Tektronix CSA7404B
348 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Tests
CSA7404B, TDS7704B, TDS7404B, TDS7254B & TDS7154B Service Manual
4-119
l. Touch Apply.
m. Verify that the instrument triggers one Unit Interval (UI) after the 1 in
the input signal (see Figure 4--37). Enter pass or fail in the test record.
4. Verify that the pattern matching circuits can do isolated 1:
a. Adjust the trigger LEVEL to trigger at 75% (+1 division) on the sine
wave.
b. Set the Format to Binary and then touch the Clear button.
NOTE. When entering the Serial Pattern Data, you only need to enter the 1 and
any trailing bits. When you touch the apply button, the instrument automatically
fills all leading bit positions with an X (don’t cares).
c. Set the Serial Pattern Data pattern bits to XXXX XXXX XXXX XXXX
XXXX XXXX XXXX XXXX XXXX XXXX XXXX XXXX XXXX
XXXX XXXX XXX1
2
.
d. Touch Apply.
e. Verify that the instrument triggers on a 1 (see Figure 4--38). Enter pass or
fail in the test record.
f. Touch the Clear button.
g. Set all Serial Pattern Data bits to X except for the nth bit, where n is the
step number. For example: if step 4c is step 1 for bit 1, the pattern for
step 2, bit 2 is XXXX XXXX XXXX XXXX XXXX XXXX XXXX
XXXX XXXX XXXX XXXX XXXX XXXX XXXX XXXX XX1X
2
.
h. Touch Apply.
i. Verify that the trigger occurs (n modulo 3) clock cycles after the 1 (see
Figure 4--38). Enter pass or fail in the test record.
j. Repeat steps g and i until all 64 bits of the Serial Pattern Data have
contained a 1.
5. Disconnect the hookup: Disconnect the cables and adapters from the inputs
and outputs.

Table of Contents

Other manuals for Tektronix CSA7404B

Related product manuals