Models 2450
-, 2460-, & 2461-EC Potentiostats And Galvanostats User's Manual Section 4:
Potential Pulse & Square Wave
077110402/November 2016 4-9
Use the Program Time end condition to stop generating the waveform after a specified time interval.
The pulse wave stops at the nearest half-cycle after the time interval elapses.
To configure the Program Time end condition:
1. Press the MENU key, then select End Conditions.
2. Select the Program Time Enabled control.
3. Select On.
4. Select the Program Time Value control.
5. Enter a value, then select OK.
Use the Source Limit end condition to stop generating the waveform if the current level reaches the
programmed source, or compliance, limit. The source limit is set to 105 percent of the selected
current measure range.
To enable the Source Limit end condition:
1. Press the MENU key, then select End Conditions.
2. Select the Source Limit Enabled control.
3. Select On.
Save or load the test parameter data
You can save your test parameters to the front-panel flash drive at any time. Up to five test parameter
configurations can be stored.
To save the test parameters:
1. From the Menu screen, select Save Settings. Note that your Menu screen options are different
depending on the test you are running.
If you have saved a settings file previously with the same File Selection number, you will be
prompted to overwrite it.
Figure 75: Selecting Save Settings