3-23
3.6.3 Sending of measurement data
Carefully certify the Patient's ID. If you send data
with mistake Patient's data (for example another
patient's ID), since the TOMEY Link (including LAN
Adapter) does not judge if they are wrong or
correct.
The TOMEY Link is able to receive a Patient's ID
through the LAN Adapter, even if the instrument is
in the state of auto power off, as long as the TOMEY
Link has been selected.
1) By pressing the TOMEY Link button $ after measure-
ment is finished, the "Patient Information, Send data
screen" (Fig. 2) is displayed. In case Auto print is avail-
able, Displays "Patient Information, Send data screen"
(Fig. 2) automatically when print out automatically.
If you pres Print out button ' before sending data, you
can print out measurement data.
2) Confirm the Patient's ID % and, if correct, press the
Send button &. Then, the Patient's ID and related mea-
surement data will be sent.
3) If Patient's ID is not correct, Press Clear button ), then
re-enter correct Patient information and measurement
again.
4) In case press the TOMEY Link button $ by mistake, if
pressing Cancel button (, return to measurement
screen with preserve Patient's ID and measurement data.
Ԙ
ԝ
ԛ
ԙ
(Fig. 1)
(Fig. 2)
'
&
(
3.6 DATA MANAGEMENT WITH "TOMEY LINK"

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tomey RC-5000 and is the answer not in the manual?

Tomey RC-5000 Specifications

General IconGeneral
BrandTomey
ModelRC-5000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals