5-5
5.5.2 Replacing of the fuse
Be sure to replace the fuse with the Power Cord
being disconnected from the receptacle, or a seri-
ous injury or struck by electricity may otherwise be
caused.
Be sure to use the fuse specified for the RC-5000.
If any operation failure occurs after the fuse has been replaced,
other troubles may have caused the failure. Then, turn the
power off immediately and report such failure to your local
distributor or representative.
1) Turn the Power Switch off.
2) Disconnect the Power Cord from the receptacle.
3) Place the coin into the slot provided in the bottom plate
of the main unit and turn it counterclockwise to remove
the fuse holder.
4) Replace the broken fuse with the new part.
5) Assemble the fuse holder in the order reverse to its re-
moving procedure.
5.5.3 Replacing of the Chin Rest Paper
1) Remove the two Chin Rest fixing pins.
2) Place the new paper on the Chin Rest and again hold the
paper to the Chin Rest with the pins.
5.5 REPLACING OF SPARE PARTS