6. Troubleshooting
6. TROUBLESHOOTING
Before you conclude your problem as a functional trouble,
check the following items. If any problem cannot be cor-
rected by the following countermeasures, consult your local
distributor or representative.
Do not remove the Main Unit Cover of this instrument. You may be
exposed to high voltage.
Do not take any corrective measures other than men-
tioned in the following items.
If your problem cannot be corrected by the following
items, ask your distributor or representative for inspec-
tion and repair.
zz
zz
z The Power Lamp and the Monitor do not light, even if the Power
switch is turned ON.
The Power Plug or the Power cord is the prob-
lem.
Check to see if the Power Plug is securely
connected to the Power receptacle. Make
sure also that the Power Cord has neither
flaw nor crack.
The Power Source is not properly supplied.
Make sure that the Power receptacle is prop-
erly supplied with electricity.
The fuse is broken.
Check to see if the fuse is proper. If broken,
replace it with the new part. (See "5.4.2 Re-
placing of the fuse".) If the replaced fuse is
again broken, the instrument may have been
disordered. If so, ask your local distributor
or representative for repair.
Cause 1
Remedy
Remedy
Cause 2
Cause 3
Remedy