8-4
8.5 EMC COMPATIBILITY DECLARATION
Caution: Any and all of medical equipment shall conform with the installing and using require-
ments stipulated by EMC (Electro Magnetic Compatibility).
Note: A portable type and a movable type RF communication device may have influence on medi-
cal instruments.
GUIDANCE AND MANUFACTURER'S DECLARATION
ON ELECTROMAGNETIC EMISSIONS
Emission test
This RC-5000 is intended for use under the following electromagnetic conditions. The customer
and the user of the RC-5000 are advised to guarantee the following use conditions of said instru-
ment.
Compatibility Elecromagnetic environment-guidance
RF emissions
CISPR 11
Group 1 The RC-5000 uses RF energy for the inter-
nal function, since RF emissions are very
low and do not induce any interference to
the internal function located close to the
electric equipment.
RF emissions
CISPR 11
Class B The RC-5000 is suitable for use in all the
clinics, and general dwelling homes and
public facilities which are directly con-
nected to public low voltage distribution
networks to supply electricity for domestic
purposes.
High frequency emis-
sions
IEC 61000-3-2
Class B
Voltage fluctuation/
flicker emissions
IEC 61000-3-3
Comparative
8.5 EMC COMPATIBILITY DECLARATION
Table 201