3-33
AutoFogging
Auto fogging which is used to reduce the accommodation
can be released temporarily. In case refractometry is difficult
because of too much nystagmus and poor fixation, make the
measurement time of refractometry shorter by releasing the
Auto Fogging temporarily.
Disable : Auto Fogging is temporarily released.
Enable : Auto Fogging works.
KRT Ø
The measurement area of keratometry can be changed tem-
porarily.
3: On the circle of 3 mm of diameter centered on the cor-
neal vertex.
3&6: On the circles of 3mm and 6 mm of diameters centered
on the corneal vertex.
6: On the circle of 6 mm of diameter centered on the cor-
neal vertex.
Print CL List
See "3.4.2 Print mode".
Eye Print
See "3.4.2 Print mode".
Kerato Print Form
See "3.4.2 Print mode".
REF Print Form
See "3.4.2 Print mode".
(Fig.3)
(Fig.4)
3.8 SETUP OF MEASUREMENT CONDITIONS