8-2
Biometry / IOL power calculations
●Measurement range
Axial length : 15.00 - 45.00mm
ACD : 1.80 - 7.00mm
Lens thickness : 2.00 - 6.00mm
●Accuracy
Measurement Accuracy : ±0.1mm
Measurement Resolution : 0.01 mm
●Factory setting of the converted velocity
Average velocity of the axial length
Phakic Eye : 1,550m/s
Dense Cataract : 1,548m/s
Aphakic : 1,532 m/s
Lens velocity
Phakic Eye : 1,641m/s
Dense Cataract : 1,629m/s
ACD velocity
Phakic Eye : 1,532 m/s
Dense Cataract : 1,532 m/s
IOL Eye : 1,532 m/s
Velocity of vitreous
Phakic Eye : 1,532 m/s
(In case of sectional velocity)
IOL Eye : 1,532 m/s
IOL velocity
Acrylic : 2,200m/s
Silicone : 1,049m/s
PMMA : 2,718m/s
User : 2,200m/s
IOL thickness
Acrylic : 0.80mm
Silicone : 1.00mm
PMMA : 0.80mm
User : 0.80mm