8-3
●IOL power calculation
Haigis standard
Haigis optimized
Hoffer
Q
Holladay 1
SRK II
SRK/T
SRK SHOWA
Shammas-PL
SRK/T Double K
●Biometry probe
Type : Solid state
Fix light : Built in the probe, Red LED
Transducer frequency : 10MHz
Tip diameter : 5.3mmφ
Dimension / Weight : 8mmφ × 100mm / 30g
Pachymetry
●Measurement range : 150 to 1,500um
●Accuracy
Measurement Accuracy : ±5μm
Measurement Resolution : 1μm
●Percent bias/plus minus bias setting range
Percent bias : 60 - 130%
Plus minus bias : -600 to +450um
●Factory setting and setting range of the converted velocity
: 1,640m/s / 1,400 to 2,000m/s
●Pachymetry probe
Type : Solid state
Transducer frequency : 20MHz
Tip diameter : 1.5mm with an angle of 45°
Dimension / Weight : 8.8mmφ × 90mm / 40g