TX9165 User Manual
TX9165-UM-EN-07 35
3.5 Sentro 8 Sensorstation Sensing Module Positions
Sentro 8 eModule and rModule
Configuration
The standard Sentro 8 Sensorstation will
accept up to four Sentro eModules in
locations A, B, C and D, plus up to four
Sentro rModules in locations E, F, G and H.
Sentro rModules cannot be fitted in locations
A, B, C and D.
Unused gas entry ports will be sealed. The
Sentro 8 Sensorstation can be adapted for
non-standard configurations according to
specification.
Checkpoint
To ensure equal internal power sharing,
do not fit Flammable Gas eModules and
Infrared eModules above one another.
Locate these eModules horizontally or
diagonally from one another.
Sensing Module Security Coding
Six coding slots at each module bay can be
individually configured using coding stops to
give a choice of 20 security codes.
Sensing Module Retainer
Sensing modules can be mechanically locked
in position for added security. Secure the
sensing module using a suitable screwdriver
and turn the retainer screw one quarter turn
clockwise.