TX9165 User Manual
TX9165-UM-EN-07 87
Checkpoint
If there is a discrepancy of greater than 5%
of reading then recalibrate the appropriate
eModule of the Sentro 8 Sensorstation.
Or consider replacing the eModule.
Checkpoint
If there is a discrepancy of greater than 5%
of reading then consider changing the gas
sensing eModule.
Checkpoint
Ensure that the appropriate Health and
Safety guidelines and applicable local
procedures are followed when handling
test gases.
The ambient temperature should be
between +20 and +30°C during the
checking procedure.
Ensure the correct operation of the
Setpoint 1 and Setpoint 2 alarms.
Institute a formal checking and
maintenance plan.
Operate a Maintenance and Calibration
Log and ensure it is updated every
time any Maintenance or Calibration
work is carried out.