Density at 2000 kg/m3
Direct Transmission (150 mm)
.25 min
1 min
4 min
Precision (kg/m
3
)
r6.8
r3.40
r1.70
Composition error (kg/m
3
)
r20.0
r20.0
r20.0
Surface error (kg/m
3
)
(100% Void)
17.0
17.0
17.0
Backscatter (98%, 100 mm)
.25 min
1 min
4 min
Precision (kg/m
3
)
r16.0
r8.00
r4.00
Composition error (kg/m
3
)
r40.0
r40.0
r40.0
Surface error (kg/m
3
)
(100% Void)
75.0
75.0
75.0
Moisture at 240 kg/m
3
.25 min
1 min
4 min
Precision (kg/m
3
)
r10.3
r5.1
r2.6
Surface error (kg/m
3
)
(1.25 mm, 100% void, kg/m
3
)
18.0
18.0
18.0
Precision is defined as rone (1) standard deviation in density
readings. This number is calculated by the ratio of the standard
deviation in the counting rate and the slope of the calibration
curve at a given density.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Troxler RoadReader 3430 and is the answer not in the manual?

Troxler RoadReader 3430 Specifications

General IconGeneral
BrandTroxler
ModelRoadReader 3430
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals