To enter the true densities now, press 〈1〉 Input True Density and
follow the gauge prompts to complete the calibration. To enter the
true density later, create a partial calibration by pressing
〈2〉 Store
Partial Calibration.
Entering the True Densities. Return to Thin Special Calibration
menu shown on page 6–18. Follow the gauge prompts to enter the
core density.
Completing Thin Special Calibration. After storing measured
density and true density, enter the desired Thin Special Calibration
name.
METHOD 2 – INPUT TRUE DENSITY FIRST. If the operator
chooses 〈2〉 Input True Density from the Thin Special Calibration
menu shown on page
6–18, and the project was a partial calibration,
the gauge displays:
Use the number keys to input the core density and press the
〈ENTER〉 key to go the next screen. If the project was not a partial
calibration, the gauge will skip the Input Core Density screen and
go directly to the following screen:
To measure the density later, create a partial calibration by pressing
〈2〉. To measure the density now, press 〈1〉. Prepare the test site as
described on page
4–6. Place the gauge on the test site and position
the source rod in the measurement position.
xxxx kg/m3
Press ENTER
1-Measure Density
2-Store Partial
Calibration