Place the reference standard block on the standard count site. As
shown in Figure 4–1, place the gauge on the reference standard
block, with the right side (keypad side) of the gauge against the
metal butt plate. Ensure that the source rod is in the standard
(SAFE) position and is firmly seated by gently tapping the handle
of the source rod.
Begin the standard count by pressing the 〈ENTER〉 key. After
taking the standard count, the gauge displays the results:
Figure 4–1. Standard Count Position
MS=xxxx
x.x%PASS
Use New Standard?