Model 3450
6–7
ENTERING TRUE MOISTURE. Under normal circumstances, the
true moisture will be entered at a time later than the measured
moisture. If measured moisture has been stored as a partial offset,
return to the Moisture Offset menu found directly above and press
2.
The gauge requests the true moisture:
COMPLETING GAUGE-DERIVED MOISTURE OFFSET.
After storing measured moisture and true moisture, the gauge will
display the values and resulting offset value, k (see formula on the
following page). The gauge offers four memory locations to store
the k value. Select the desired memory location to save the value.
Edit a Partial Moisture Offset
If the operator saved a partial offset, then the gauge will display a P
beside the option entered before. Select the value to be completed.
If a value already entered is selected, then the gauge prompts the
operator to overwrite. To overwrite the partial offset, press the
YES key. To use the stored partial offset, press the NO key and
select the other value.
To take measurements, see page 6–6. To enter the true moisture, see
the description above.
Input True Moisture
xxxxx
%
ENTER When Done
6. OFFSETS & CALIBRATIONS