The gauge displays:
Begin the measurement by pressing the 〈START〉 key. The gauge
displays the progress. Then, it displays the reading. Press the
〈ENTER〉 key.
To enter the true density now, press 〈1〉 and follow the gauge
prompts to complete the calibration. To enter the true density later,
create a partial calibration by pressing
〈2〉.
ENTERING TRUE DENSITY. If the measured density has been
stored as a partial calibration, return to Special Calibration menu as
described on page 6–11. The gauge requests the true density:
Use the number keys to enter the true density. Press the 〈ENTER〉
key.
COMPLETING SPECIAL CALIBRATION. After storing the
measured density and true density, enter the desired Special
Calibration name.
Depth: xxx in
Time 4 min
Press START
1-Input True Density
2-Store Partial
Calibration
Density
xxxx kg/m3
Press ENTER