Model 3450
6–15
Edit a Partial Special Calibration
If a partial calibration was saved earlier, the gauge will display a P
beside the option entered before. Select the value to be completed.
If the value already entered is selected, the gauge prompts the
operator to overwrite. To overwrite the partial calibration, press the
YES key and go to page 6–13, Create a Special Calibration. To
use the stored partial calibration, press the
NO key and select the
other value.
To take measurements, see page 6–13. To enter the true density, see
page 6–14.
Erase Special Calibrations
At the Special Calibration menu (page 6–11),
press 5. The gauge
displays the Erase Special Calibration menu:
From the Erase Special Calibration menu, the operator can erase
either a single special calibration or all special calibrations.
To erase a single special calibration, use the arrow keys to scroll
through the special calibrations names displayed on the first line.
When the gauge displays the special calibration to be erased, press
1. At the Are you sure . . . display, press the YES key. The
gauge erases the special calibration and returns to the Ready screen.
To erase all special calibrations, press 2 at the Erase Special
Calibrations menu. At the Are you sure . . . display, press the
YES key. The gauge erases all the special calibrations and returns
to the Ready screen.
SP CAL: Your Special
Calib
1
Erase This Sp Cal
2
Erase All
Sp
Cal
6. OFFSETS & CALIBRATIONS