Page 89
Message /
Symbol /
Problem
Failure Description
Corrective Action
Canceled
By User
This message is displayed
after the START/STOP key
is pressed and cycle
aborted.
Wait until "cycle failed
canceled by user" or
"cycle end canceled
by user" is displayed.
Perform a new cycle.
Cycle Failed
This message and symbol
are displayed if an error
occurs before sterilization
cycle is completed.
Perform a new cycle.
Air Error
This message is displayed
at the end of the cycle If
the autoclave does not
reach the atmospheric
pressure after 10 minutes.
Wait until the
autoclave reaches the
atmospheric pressure
and perform a new
cycle.
Compresse
d air supply
error
(ELVCPVG-
D only)
This message is displayed
in case of a compressed air
supply malfunction.
Check and fix the
compressed air supply.
Periodical
check time
exceeded -
Please call
for service
The periodical maintenance
time has passed.
Call for service.
Cycle
counter
exceeded -
Please call
for service
Number of cycles, since
last periodical
maintenance, exceeded the
preset number as defined
by "cycle counter"
parameter.
Call for service.
Power
Down
This message is displayed
if power down has occurred
during the cycle. (this
message will print out in
the printer after the
autoclave will turn on).
Turn on the autoclave
and wait until the
autoclave is ready
(reaches the safe
condition) and perform
a new cycle.
Supply
distilled
water error
(digitat input
option)
This message is displayed
in case of a mineral free
water supply malfunction.
Check and fix the
mineral free water
supply
Supply
water error
(digitat input
option)
This message is displayed
in case of a city (tap) water
supply malfunction.
Check and fix the city
(tap) water supply.
Compresse
d air supply
error (digitat
This message is displayed
in case of a compressed air
supply malfunction.
Check and fix the air
supply.