Page 90
Message /
Symbol /
Problem
Failure Description
Corrective Action
input option)
No Water
This message is displayed
if the electrode in the
chamber did not sense
water within the preset
time.
1. check and fix
the mineral free water
supply.
2.check and clean the
water inlet filter.
3. Clean the water
level electrode.
Analog
Input Error
This message is displayed
when any Temperature
sensor or Pressure sensor
is disconnected or out of
range.
Call for service.
Chamber
temperature
not in range
This message is displayed if
the temperature in the
chamber is too high or too
low from the normal range.
Call for service.
Chamber
pressure not
in range
This message is displayed if
the pressure in the chamber
is too high or too low from
the normal range.
Call for service.
I/O Card
Failed
This message is displayed
if I/O card is faulty (both
while cycle is running or
not).
Call for service.
I/O card is
not
connected
This message is displayed
if I/O card is disconnected
(both while cycle is running
or not).
Call for service.