This message is displayed when any
Temperature sensor or Pressure sensor
is disconnected or out of range.
Chamber
temperature not
in range
This message is displayed if the
temperature in the chamber is too high
or too low from the normal range.
Chamber
pressure not in
range
This message is displayed if the
pressure in the chamber is too high or
too low from the normal range.
This message is displayed if I/O
card is faulty (both while cycle is
running or not).
This message is displayed if I/O
card is disconnected (both while
cycle is running or not).
This message is displayed if the
temperature drops for more than 1
second below the sterilization
temperature during sterilization cycle.
This message is displayed if the
temperature raises 7°F (4°C) above
sterilization temperature during the
sterilization stage for 2 seconds during
sterilization cycle.
This message is displayed if the
system cannot reach the required
temperature, in the chamber, within 10
minutes.
This message is displayed if the
system cannot reach the required
temperature, in the chamber, within the
preset time.
Verify that the autoclave is not
overloaded.
This message is displayed if the
system cannot reach the required
temperature, in the chamber, during
the optional "Keep Heat" stage, within
the preset time.
Verify that the autoclave is not
overloaded.
This message is displayed if Chamber
Pressure drops below the sterilization
pressure (134°C = 304 kPa, 121°C =
205 kPa) for 2 seconds during the
sterilization stage.
This message is displayed if Chamber
Pressure raises 4.2 psi-29 kPa above