sterilization pressure (134°C = 304
kPa, 121°C = 205 kPa) for 2 seconds
during the sterilization stage.
This message is displayed if the system
cannot reach atmospheric pressure ±
5kPa during the ending stage.
This message is displayed if the
system cannot reach preset pressure
within 10 minutes from the beginning
of the exhaust stage.
This message is displayed if the
system cannot reach the required
pressure conditions in the chamber,
after preset time, during the air
removal stage.
Verify that the autoclave is not
overloaded.
RTC Error -
Please Set
Current Date
and Time
This message is displayed in order
to set the date and the time.
Set Current Date And Time.
If the problem persists, call the
technician.
This message is displayed if the real
time clock is faulty.
Door is open
(During the
cycle)
This message is displayed when the
door is open: During the cycle.
Close the door to perform a
new cycle.
This message is displayed after the
START/STOP key is pressed and
cycle aborted.
Wait until "cycle failed –
canceled by user" or "cycle
end – canceled by user" is
displayed.
Perform a new cycle.
This message and symbol are
displayed if an error occurs before
sterilization cycle is completed.
This message is displayed at the end
of the cycle If the autoclave does not
reach the atmospheric pressure after
10 minutes.
Wait until the autoclave
reaches the atmospheric
pressure and perform a new
cycle.
Compressed
air supply
error
This message is displayed in case of
a compressed air supply
malfunction.
Check and fix the compressed
air supply.
Periodical
check time
exceeded -
Please call for
service
The periodical maintenance time has
passed.
Cycle counter
exceeded -
Please call for
Number of cycles, since last
periodical maintenance, exceeded
the preset number as defined by