"cycle counter" parameter.
This message is displayed if power
down has occurred during the cycle.
(this message will print out in the
printer after the autoclave will turn
on).
Turn on the autoclave and
wait until the autoclave is
ready (reaches the safe
condition) and perform a new
cycle.
Supply
distilled water
error (digitat
input option)
This message is displayed in case of
a mineral free water supply
malfunction.
Check and fix the mineral
free water supply
Supply water
error (digitat
input option)
This message is displayed in case of
a city (tap) water supply
malfunction.
Check and fix the city (tap)
water supply.
Compressed
air supply
error (digitat
input option)
This message is displayed in case of
a compressed air supply
malfunction.
Check and fix the air supply.
This message is displayed if the
electrode in the chamber did not
sense water within the preset time.
1. check and fix the mineral
free water supply.
2.check and clean the water
inlet filter.
3. Clean the water level
electrode.