Table 4-1 List of user and advanced level commands. Advanced level commands are marked bold.
1st (TOP)
LEVEL
2nd LEVEL 3rd LEVEL 4th LEVEL 5th LEVEL DESCRIPTION
CLOSE Release port for message transmission, i.e. abort command dialog
BACK Back to normal user command set
RESET NO Do not reset
YES Make full reset
SET
ALGORITHM DEFAULTS Set default algorithm parameter settings
CONTROL BLOWER ON Turn window conditioner blower ON
OFF Turn window conditioner blower OFF
BLOWER MANUAL Set window conditioner blower to manual control
AUTO Set window conditioner blower to automatic control
INHEATER ON Turn internal heater ON
OFF Turn internal heater OFF
OUTHEATER ON Turn window conditioner heater ON
OFF Turn window conditioner heater OFF
DATA_ACQ
AUTOADJ ON Data Acquisition parameters are software controlled
OFF Data Acquisition parameters are manual controlled
COMP COARSE Set internal crosstalk compensation setting coarse code value
FINE Set internal crosstalk compensation setting fine code value
INTERVAL 15...120 Data Acquisition: Set interval for measurement and message sending.
TRANSMIT POWER_OF_P Set pulse energy input code value
FACTORY INLASER Set pulse energy input code value (at start-up)
OUTLASER Set pulse energy target value for software adjustment
RECVALUE Set receiver test reference value
WIN_CLEAN Set clean window reference value