EasyManua.ls Logo

Vaisala CT25K - 4.6 Prevailing Parameter Settings

Vaisala CT25K
139 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Ceilometer CT25K
User’s Guide CT25K-U059en-2.1
32
1st (TOP)
LEVEL
2nd LEVEL 3rd LEVEL 4th LEVEL 5th LEVEL DESCRIPTION
GET VALUE TEMPERATURE BLOWER Print value of measured blower temperature, units: °C
CPU Print value of measured CPU board temperature, units: °C
LASER Print value of measured laser temperature, units: °C
LENS Print value of measured temperature adjacent to lens, units: °C
OUTSIDE Print value of measured outside temperature, units: °C
VOLTAGE BCIRCUIT Print status of battery circuit, connected / disconnected = 1/0
BATTERY Print value of battery voltage, approx. +13V
CHARGE Print value of battery charge voltage, approx. +13V
VCA Print value of internal raw voltage, approx. +25V
PHV Print value of Receiver high voltage, approx. +200V
PFB Print value of Receiver switcher internal feedback voltage, appr +2V
VALUE VOLTAGE P65 Print value of Transmitter high voltage, approx. +65V
P18 Print value of general internal supply voltage +18V
P13 Print value of Receiver supply voltage +13V
P12 Print value of general internal supply voltage +12.5V
P5G Print value of general internal supply voltage +5V
P5R Print value of Receiver supply voltage +5V
M13 Print value of Receiver supply voltage -13V
M12 Print value of general internal supply voltage -12.5V
M5R Print value of Receiver supply voltage -5V
M5G Print value of general internal supply voltage -5V
VERSION Print identifier for Software version
VLIM Print sky condition vertical visibility reporting limit (%)
START Start single-cycle measurement in operation mode STANDBY. Refer
to command SET OPER_MODE STANDBY.
STOP Stop internal compensation routine