3.5 Duct Installation Kit
The duct installation kit includes a plastic pipe with a flange (Vaisala order code: 215619). To
install the HMP110 probe with the duct installation kit, drill a hole to the duct wall, assemble the
probe to the duct installation kit, slide the probe head through the hole, and attach the flange
to the duct wall with four screws.
Figure 6 Probe Installation with the Duct Installation Kit
1
Tension screw
The installation depth (distance L) can be adjusted and locked in place with the tension screw.
Chapter 3 – Installation