Syntax Description
Example:
vers
HMT120 / 0.1.0.103
4.7 Serial Line Output and Communication
Table 14 R Command
Syntax Description
r<cr>
S
Use the r command to start the continuous
outputting of measurement values as an ASCII
text string to the serial line. The output always
includes the readings of the currently selected
analog output quantities.
Outputting the results continues in intervals
issued with the command intv. You can stop the
output by entering the s command.
You can
define the output data formatting with
the form command. With X parameter the
transmitter will output RH + T value in
predefined format regardless of what quantities
are selected.
Example:
r
RH= 25.10% T= 24.77'C
RH= 25.12% T= 24.96'C
...
Table 15 S Command
Syntax Description
s<cr>
Stop the continuous outputting that was started
with the r command.
HMT120 User Guide M211244EN-C