6.3.2 1-Point Adjustment Using a Calibrator
Complete the steps listed in Adjusting RH and T with MI70 Hand-Held Indicator
(page 56) before continuing with this adjustment option.
1. Remove the filter from the transmitter's probe and insert the probe head into the
reference condition.
2. Press ADJUST to continue adjusting.
3. Choose 1-point adjustment from the MI70 adjustment menu and press SELECT.
4. Press READY when the reading has stabilized in the reference condition (can take 30
minutes or more). You can follow the stabilization from the GRAPH display.
5. Enter the correct reference value with the arrow buttons and press OK.
6. Confirm the adjustment by pressing YES.
7. The adjustment is done. Press BACK and EXIT to return to the basic display.
8. Turn
o the MI70 and detach the connection cable.
6.3.3 2-Point Adjustment Using a Calibrator
Complete the steps listed in Adjusting RH and T with MI70 Hand-Held Indicator
(page 56) before continuing with this adjustment option.
HMT120 User Guide M211244EN-C
58