5.4 Replacing the HUMICAP Sensor
To replace the HUMICAPâ sensor on a fixed or remote probe:
1. Remove the filter (plastic grid or sintered stainless steel).
2. Remove the damaged sensor and insert a new one.
3. Recalibrate the probe.
4. If the
filter is dirty, replace it to ensure maximum lifetime and fast sensor response. Do not
attempt to clean the filter.