Syntax Description
Example (no active errors):
>aout
*** ANALOG OUTPUT 1 ***
Ch1 Analog out mode : 4_20MA
Ch1 Notification : ON
Ch1 Error level : 3.6
Ch1 Status : ON
Ch1 Quantity : RH
RH lo : 0
RH hi : 100
RH : 23.32 %
Current : 7.73 mA
*** ANALOG OUTPUT 2 ***
Ch2 Analog out mode : 4_20MA
Ch2 Notification : ON
Ch2 Error level : 3.6
Ch2 Status : ON
Ch2 Quantity : T
T lo : -60
T hi : 100
T : 23.66 'C
Current : 12.37 mA
>
Table 12 System Command
Syntax Description
system<cr>
Show probe firmware information.
Example:
system
Device Name : HMT120
SW Name : HMT120
SW model : HMT120
SW version : 1.0.0.500
Serial number : A1234567
>
Table 13 Vers Command
Syntax Description
vers<cr>
Show firmware version of the probe.