Syntax Description
Example (set a new information text):
ctext Calibration lab 2
Calib. info: "Calibration lab 2"
>
Table 11 Aout Command
Syntax Description
aout<cr>
The aout command shows the following
information:
Analog out mode is the output mode for the
channel, for example 0 ... 10 V.
Error level is the level that the analog output
is set to in case of transmitter error.
Status indicates the current status of the
channel. The statuses are:
ON: Normal measurement operation.
OFF: No quantity selected for the channel
using the asel command. See Table 28
(page 40).
ERROR: Transmitter error, channel set to
error level.
TEST: Analog channel being tested using
the atest command. See Table 29
(page 42).
HMT120 User Guide M211244EN-C