Syntax Description
Example 2: 2-point calibration
>ct
T 22.0007 Ref1 ? 22
Press ENTER to continue or ESC to exit
T 44.9847 Ref2 ? 45
OK
>
Example 3: No calibration, only updating the value
>ct
T 22.0007 Ref1 ?
T 22.0145 Ref1 ?
Calibration terminated with ESC
>
Table 21 CTCLR Command
Syntax Description
ctclr<cr>
Use the ctclr command to restore the probe
temperature factory calibration.
Example:
>ctclr
OK
>
Table 22 ACAL Command
Syntax Description
acal<cr>
Use the acal command to calibrate the analog
current loop outputs.
After you give this command, the CH1 analog
output of the transmitter is set to the low limit
(4mA). Measure the output and enter the
measured value. After entering the low limit, you
must measure and enter the high limit (20mA)
also. The same procedure is then repeated with
CH2. Calculated coecients are printed after the
measured values are given.