6.4.1 Connections and Selecting the Calibrator Function
1. To select the calibrator function from the HMI41, press the ON/OFF button until you can
see some text on the display. Then release the ON/OFF button and within 1 ... 2 seconds,
press both ENTER and MODE buttons until the following display appears:
After a few seconds, the display changes to show the following:
If the basic settings (display units, automatic power o function, display
quantities, and pressure) have to be changed, see the HMI41 documentation
set.
2. Press ENTER repeatedly until the following display appears:
3. To calibrate the transmitter, select number 3 with buttons (number up) and
(number down) and then press ENTER.
4. Next the baud rate appears on the display. Select the baud rate 19200 (= 19.2 on the
HMI41 display) and press ENTER. Configure the rate with the and buttons if
necessary.
5. The serial communications selection appears. Select the following settings: N, 8, 1 and
press ENTER.
Configure the settings with the and buttons if necessary.
HMT120 User Guide M211244EN-C
62