Responding to System Alarms
7-12 LigaSure Vessel Sealing Generator Service Manual
014
015
016
017
018
019
020
021
022
023
024
INVALID_BIP_OPS
Invalid value in available operations
variable read back from smart connector
table.
INVALID_VS_OPS
Invalid value in available operations
variable read back from smart connector
table.
FB_ILLEGAL_STATE_TRANSITION
The combination of the new state
requested by the master microprocessor
and the current state of the feedback
microprocessor represent an illegal state
transition request.
FB_ILLEGAL_STATE_VALUE
An illegal state value was detected in the
feedback state transition checking
function.
FB_ILLEGAL_POWER_DEACTIVATIO
N_STATE_VALUE
An illegal state value was detected in the
feedback’s power deactivation state
machine state variable.
FB_INVALID_MODE_CHANGE
An attempt was made to transition
directly from one RF output mode to
another without first disabling RF (from
master)
FB_INVALID_POWER_CHANGE
An attempt was made to change power
output while SEAL output was activated.
FB_INVALID_MODE_COMMAND
RF output mode (function) sent from the
master to the feedback was invalid.
FB_INVALID_CAL_DATA
Calibration data copied from DPRAM to
local RAM has an invalid CRC and is,
therefore, unusable.
FB_INVALID_FEEDBACK_MODE
Invalid feedback mode state variable
value.
MASTER_FPANEL_TEST_
TIMEOUT
Master failed to signal feedback that
front panel testing has been completed
in time.
Turn the power switch off (O) then on (|) again.
If the error reappears, replace the control board.
Number Description Recommended Action