Calibrating the LigaSure Generator
6-16 LigaSure Vessel Sealing Generator Service Manual
Calibrating the LigaSure Generator
Calibration comprises seven steps. During calibration you verify information
specific to the LigaSure generator, adjust the date, and adjust the clock. You also
adjust several values, or factors, that ensure the proper operation of the generator.
The following table summarizes the calibration steps and the values you can
adjust. Certain values you cannot adjust, but you must verify them.
Notice
After completing any calibration step, follow the instructions displayed on the
terminal.
Step Description Adjustable?
1 LigaSure Generator Data
Generator Model Number
Master microcontroller software version
Feedback microcontroller software version
No (verify value)
No (verify value)
No (verify value)
2 Calendar
Month
Day of the month
Year
Yes
Yes
Yes
3 Clock
Hour
Minute
Yes
Yes
4 Current sense gain factor Yes
5 Voltage sense gain factor Yes
6 ECON factor Yes
7 Save calibration values Yes