Standards and IEC Classifications
4-8 LigaSure Vessel Sealing Generator Service Manual
220-240 VAC
Cable - H05VVF3G1.0 VDE, maximum length 15’ (5 meters)
Plug - minimum 6 A - 250VAC
Unit receptacle - IEC female, minimum 6 A - 250VAC
Standards and IEC Classifications
The LigaSure Generator meets all pertinent clauses of IEC 60601-1 second
edition and IEC 60601-2-2 third edition.
Class I Equipment (IEC 60601-1)
Accessible conductive parts cannot become live in the event of a basic insulation
failure because of the way in which they connect to the protective earth
conductor.
ATTENTION
Consult accompanying documents.
The generator output is floating (isolated) with respect to
ground.
DANGER
Explosion risk if used with flammable anesthetics.
To reduce the risk of electric shock, do not remove the cover.
Refer servicing to qualified service personnel.
Non-Ionizing Radiation
Classified with respect to electrical shock, fire, and
mechanical hazards only in accordance with UL60601-1 and
CAN/CSA C22.2 No. 601.1.
F