Output Characteristics
4-14 LigaSure Vessel Sealing Generator Service Manual
Output Characteristics
Maximum Generator Output
* Open circuit values obtained with a Pearson 411 current measurement with a 1K
ohm load attached to the handset.
** Crest factor is an indication of a waveform’s ability to coagulate without a
cutting effect.
Mode
Maximum Open
Circuit Voltage
V
pp
(V
p
)
Maximum Short
Circuit Current*
A
rms
Maximum Power
Setting
watts
Crest
Factor**
Macrobipolar 760 (380) 2.2 95 1.5
Bipolar 335 (168) 2.2 95 1.5
Seal 575 (288) 4.4 150 1.5