Performance Characteristics
Technical Specifications
LigaSure Vessel Sealing Generator Service Manual 4-3
Transport and Storage
Duty Cycle
Under maximum output settings and rated load conditions (100 ohm load) the
generator is suitable for activation times of 10 seconds on, 30 seconds off, for one
hour.
Internal Memory
Audio Volume
The stated audio levels are at a distance of one meter. Alarm tones meet the
requirements of IEC 60601-2-2.
Ambient temperature
range
– 34° to 70° C (– 29° to 158° F)
Relative humidity 0% to 95%, noncondensing
Atmospheric pressure 500 to 1060 millibars
Duration of storage If you stored the generator over one year, check the
battery to measure the Vdc minimum and complete a
full checkout (including calibration) before use.
Contact Valleylab Service for information.
Memory type Nonvolatile, battery-backed RAM
Battery type 3 V lithium button cell
Battery life 5 years