EasyManua.ls Logo

Valleylab LigaSure - Low Frequency Leakage Current; High Frequency Leakage Current

Valleylab LigaSure
168 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Characteristics
4-6 LigaSure Vessel Sealing Generator Service Manual
Low Frequency (50-60 Hz) Leakage Current
(AAMI HF-18-1993)
High Frequency (RF) Leakage Current (IEC 60601-2-2)
Enclosure source
current, ground open
< 300 µA
Source current, patient leads, all outputs:
Normal polarity, intact
ground
< 10 µA
Normal polarity,
ground open
< 50 µA
Reverse polarity,
ground open
< 50 µA
Sink current at high
line, all inputs
< 10 µA @ 120 V
< 50 µA @ 264 V
Bipolar RF leakage
current
69 mA rms
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com