Output Characteristics
4-16 LigaSure Vessel Sealing Generator Service Manual
F i gu r e 4 -2 .
Output voltage at specific
intensity settings in the
macrobipolar and bipolar modes
Table 4-2.
Open Circuit Output Volts Peak
(rms)
Mode Generator Power Setting — Watts
10 25 50 75 90 95
Macro Bipolar 133 (94) 201 (142) 293 (207) 355 (251) 354 (250) 287 (203)
Bipolar 51 (36) 82 (58) 115 (81) 139 (98) 151 (107) 266 (110)