Inspecting the Generator and Accessories
Setup, Tests, and Adjustments
LigaSure Vessel Sealing Generator Service Manual 6-7
5. Check the vessel sealing handset receptacle on the front panel for obstructions
or damage. Connect a footswitching vessel sealing instrument to this
receptacle, inserting the handset plug firmly into the receptacle to verify a
secure fit.
If the connection is loose, re-apply the handset to secure the handset/
receptacle mating. If the connection is loose, try another handset. If the
connection is still loose, replace the front panel. Refer to Chapter 8, Front
Panel and Display Board for further instructions.
6. With the bipolar and vessel sealing handsets properly mated, check that the
smart indicator lamp illuminates green. If the smart indicator lamp illuminates
red, use the following procedure:
a. Verify that you have connected the correct LigaSure handset to the seal
receptacle.
b. Verify that the LigaSure handset fits securely in the seal receptacle.
c. Verify that you have connected a smart bipolar adapter to the bipolar
receptacle.
Power Cord
7. Unplug the power cord from the wall receptacle, then remove the cord from
the unit.
8. Inspect the power cord for damage.
9. Reconnect the power cord to the generator and wall receptacle.