2920 Float & Tape Transmitter 5 Display and Configuration Interface
90 Installation and Operations Manual
Display Only Description
Avg Temp Average temperature
Vap Temp Average vapor temperature
Elem #00 Temperature conversion value of internal precision
100
Ohm resistor.
Elem #01 through #16 Temperature measured by element #1 thrrough
element #16
Elem #17 Temperature conversion value of internal precision
200
Ohm resistor.
Meas Level The liquid level used for the average temperature
calculations, it is automatically updated by the 2920 to
be the actual measured level.
Status Status code of the NMT.
Diag Code When the NMT electronics detect an error this
parameter shows the error code detected.
Univ Cmd Rev Revision of the universal command set the NMT
supports.
Dev Cmd Rev Revision of the device commands that the NMT
supports.
Softw. Version The software version of the NMT.
Hardw. Version The hardware version of the NMT.
Cmd Status The status of command, as follows:
•Start
•Complete
•Error
Cmd Error The number of error if Command fails.
Comm. Tries The number of times the 2920 has polled the NMT.
Comm. Fails The number of time the communications from the 2920
to the NMT has failed.
Elapse Time of last update.
Pnt Status Byte value indicating status of point. Values are:
0: No error
Pnt CheckSum CRC-16 Checksum of point's static configuration
parameters.
Command
Level The measure level sent from the 2920 to the NMT.
Command The current command sent to the NMT.
Scan Cmd The command used to set the type of scanning used by
the 2920 - Controlled by the application.