6.1.5. 1-Point Bright Calibration
To perform a 1-point bright calibration.
1. Run the VAKAP.exe file.
Norm C:\VCaptureSW\Acquisition\CEPH\WideCEPHG\Norm
Fast C:\VCaptureSW\Acquisition\CEPH\WideCEPHG\Fast
2. On the dialog box, Select Settings and follow the steps to set exposure conditions:
1) Go to [BrightCalibration] and enter 1 for the NumbPoints.
2) Check if the tube voltage (kVp) and current (mA) are set as the table below. If
necessary, type each value according to the table:
Exposure Condition
Values
kVp0 660
mA0 900
kVp1 650
mA1 1000
3) Click Save.