Page 93 of 244
2.5.4.3 Pin Center Alignment (X-axis:95)
Follow the instructions below to align the pin center:
1. Unscrew three bolts (1).
2. Turn the side bolt (2) to move the chinrest handle right or left.
• When moving the handle right, the pin center moves to the right.
• When moving the handle left the pin center moves to the left.
3. After the pin center alignment, recover the value by sending the original HFST value.
ex. If it is [SPM_HFST_3800], input [SPM_HFST_200].
When moving to the left,
the pin moves to the left
When moving to the right,
the pin moves to the right