2.5.4.4 Rotator Y, Z -axis setting
1. Put the chin block and the normal bite on the unit as shown below.
2. Run VAKCAP.exe file.
3. Click the Capture/Recon tab and send the [SPM_ARCT] command to initiate the system
into CT mode.
4. When the equipment stops operating, send the following commands in the table below to
supply power to the CBCT Sensor.
Command
Function
[SPM_CSON]
CBCT sensor Power On
[SPM_FISS_0001]
Sensor Sensitivity & Binning mode setting
//0000: High Gain, 2x2 Binning
//0001: High Gain, 4x4 Binning
//0002: Low Gain, 2x2 Binning (default)
//0003: Low Gain, 4x4 Binning
[SPM_FIPM_0001]
Sensor Partial mode setting
// Sensor size: 0000~0032
[SPM_FISM_0001]
Sensor Internal/External mode setting
//0000: Internal
//0001: External
[SPM_COLM_0003]
Collimator Open position
If Collimator is not completely opened, adjust with
[SPM_CDST_XXXX]
5. Send the [SPM_ABMS] command to acquire dark.
6. Click the Soft Calib tab and click the Dark button to acquire Dark Calibration Data.