Page 94 of 244
2.5.4.4 Rotator Y, Z -axis setting
1. Put the chin block and the normal bite on the unit as shown below.
2. Run VAKCAP.exe file.
3. Click the Capture/Recon tab and send the [SPM_ARCT] command to initiate the system
into CT mode.
4. When the equipment stops operating, send the following commands in the table below to
supply power to the CBCT Sensor.
[SPM_CSON]
CBCT sensor Power On
[SPM_FISS_0001]
Sensor Sensitivity & Binning mode setting
//0000: High Gain, 2x2 Binning
//0001: High Gain, 4x4 Binning
//0002: Low Gain, 2x2 Binning (default)
//0003: Low Gain, 4x4 Binning
[SPM_FIPM_0001]
Sensor Partial mode setting
// Sensor size: 0000~0032
[SPM_FISM_0001]
Sensor Internal/External mode setting
//0000: Internal
//0001: External
[SPM_COLM_0003]
Collimator Open position
If Collimator is not completely opened, adjust with
[SPM_CDST_XXXX]
5. Send the [SPM_ABMS] command to acquire dark.
6. Click the Soft Calib tab and click the Dark button to acquire Dark Calibration Data.