VTK‐VS‐BA‐004e‐B|UserManualMDS3200A | Communication Interfaces
40
1. *ESR?
Description: Thiscommandshowsthelast10errorcodescreatedbythesystem,
startingwiththelatestevent.Iftheresponseconsistsoftentimes‚0‘,
noerrormessageispresent.
Example:
Input: *ESR?
Result: Listofthelast10errormessages
Return: 8AUTORANGEERROR
9WATCHDOGTIMEOUT
0
0
0
0
0
0
0
0
2. *IDN?
Description: Devicespecificinformation,formattedasfollows:manufacturer,model,
softwareversion.ImportantforcontactwithourTechnicalSupport.
Example:
Input: *IDN?
Result: VERMES,MDV3200A,065.411.001
Return: VERMES,MDV3200A,065.411.001
3. ADJUST:?
Description: Thisinstructionisusedtoverifythesuccessoftheadjust.Theresult
consistsinoneofthefollowingoptions:
Result:0
UnknownState.Thesystemisnotabletointerpretthequestionand
hastobeswitchedOFFandONagain.Theadjustmustberepeated.This
event
howeverisanexception.
Result:1
Thepositionofthenozzleunitistoolowandmustberaisedonthefluid
box.Anadjustisrequired(ormustberepeated).
Result:2
ThepositionofthenozzleinsertrelativelytothetappetisOK.
Result:3
Thenozzleunit
onthefluidboxmustbescrewedindownwarddirec‐
tion,asthecurrentpositionistoohigh.Theadjustmustbeperformed
(orrepeated).
Example:
Input: ADJUST:?
Result: Thesysteminformsaboutthecurrentnozzleposition.