VTKVSBA004eB|UserManualMDS3200A | Error Messages
65
14 WatchdogTimeOutPressEnter
Abnormalend(crash)ofthecontrolunit.
ErrorCodeDisplay: EEPROMErrorPressEnter
ErrorCodeStatusMenu: WATCHDOGTIMEOUT
CorrectiveAction:
Themessagehastobeacknowledgedby[enter].
Thecontrolunitrestarts.
15 ERROR999
Theproblemcannotbeassignedtoanyerrorothercodeofthelist.
ErrorCodeDisplay: ERROR999
ErrorCodeStatusMenu: ERROR999
CorrectiveAction:
SwitchOFF thecontrolunit,andswitchitonagainafterca.
15sec.Themessageisreportedintheerrorlist,withno
consequencestocontrolunitorvalve.
16 CounterErrorPressEnter
Duringstartupofthecontrolunit,thesystemreadsthecounters(e.g.Nozzlecounter,Tap
petcounter).Ifthisproblemarises,thecorrespondingvaluesinfuturemaybedoubtful.
ErrorCodeDisplay: CounterErrorPressEnter
ErrorCodeStatusMenu: Nomessage
CorrectiveAction:
Acknowledgeby[enter].Acorrespondingremarkaboutthis
eventinthemaintenancescheduleisrecommended.
17 NOHEATERPressEnter
EventhoughtheheateroptionhasbeenactivatedintheHeaterMenu,thiselementisnot
detected.
ErrorCodeDisplay: NOHEATERPressEnter
ErrorCodeStatusMenu: Nomessage
CorrectiveAction:
Themessagehastobeacknowledgedbypressing[enter].
‐Heaterstatusin“Heater“menuissetto“OFF“.
Inspecttheheatercableconnection.Worn,kinkedor
charredcablesmustbeexchanged.
Heaterdefectchangetheheater
18 USARTBufferOverflow
This problem sometimes occurs during data transmission via serial interface. The buffer is
fullandthereforenotabletohandleinflowingdata.Thereforethemessage“USARTBuffer
Overflow”isreturnedviaserialinterface. LEDsarenotlit.
ErrorCodeDisplay: Nomessage
ErrorCodeStatusMenu: Nocode
ErrorCodeMonitor: USARTBufferOverflow
CorrectiveAction:
Datatransmissionmustbeinterrupted
Senddataagain